Where optimum product protection is required, The Walker Vertical Downflow system is the UK’s leading clean air solution for nuclear medicine bench top procedures. Its key features are:
- Tested to BS EN ISO 14644-1 Class 5.
- Spacious, open working area.
- Available in various sizes .
- Clear side screens for maximum visibility.
- Low energy high efficiency fans.
- Protects the product from airborne contaminants at all times.
- Recirculating airflows manage the lab environment cleanliness too.
Apart from that standard range of products we also offer a bespoke design and build capability, this allows you to overcome any issues with space, process or equipment conflicts.
Controls:
- Key switch for the on/off function.
- Socket on/off push switch.
- Alarm Mute switch.
- Audible and Visual Low Airflow Alarm.
Optional Extras:
- Additional Power outlets.
- Front screens.
- Network sockets.
- UV Lamps.
- Front Closure Panels.
Details | BT9 | BT10 | BT12 | BT15 | BT18 | |
---|---|---|---|---|---|---|
Width (DIM A) | mm | 900 | 1000 | 1200 | 1500 | 1800 |
Depth | mm | 675 | 675 | 675 | 675 | 675 |
Height | mm | 1230 | 1230 | 1230 | 1230 | 1230 |
Air Flow (at 0.45m/s) | m³/s | 0.28 | 0.32 | 0.38 | 0.47 | 0.57 |
Load | Watts | 120 | 130 | 130 | 180 | 235 |
Filter | HEPA H14 99.995% High efficiency | |||||
Lighting | <750 Lux at working area | |||||
Noise | 57 to 62 dBA |